Stability Climate Chamber ICH750L
Temperature
Working temperature range | from 10°C up to 60°C with humidity |
---|---|
Setting accuracy temperature | 0.1°C |
Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity
Humidity | humidity supply with distilled water from external tank by self-priming pump |
---|---|
Humidification | humidification by hot steam generator |
Humidity | active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 % |
Control of standard components
Light | daylight and UV-light can be programmed individually via controller |
---|---|
Light | cold white light /- 6 %, for UV-light /- 10 % |
Light | illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable values each, temperature/humidity, for two freely selectable values CO2 |
adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime |
Ventilation
Fan | forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps |
---|
Communication
Documentation | programme stored in case of power failure |
---|---|
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
---|---|
Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY | additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
Alarm | visual and acoustic |
Heating concept
Air jacket | air jacket heating system for gentle all-around heating |
---|---|
Heating concept | no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior |
Cooling | energy-saving use of CFC-free cooling/heating system (refrigerant R134a) |
Defrosting | highly efficient automatic defrosting system |
Standard equipment
Internals | 2 stainless steel grid(s), electropolished |
---|---|
Works calibration certificate | standard values 10 °C and 37 °C, 60 % rh at 30 °C |
Door | inner glass doors |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Stainless steel interior
Volume | 749 l |
---|---|
Dimensions W x H x D in mm | w(A) x h(B) x d(C): 1040 x 1200 x 600 mm |
Max. number of internals | 14 |
Max. loading of chamber | 200 kg |
Max. loading per internal | 30 kg |
Textured stainless steel casing
Installation | on lockable castors |
---|---|
Dimensions | w(D) x h(E) x d(F): 1224 x 1950 x 834 mm |
Housing | rear zinc-plated steel |
Electrical data
Voltage Electrical load | 230 V, 50/60 Hz approx. 1350 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 1350 W |
Ambient conditions
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
---|---|
Ambient temperature | 15 °C to 28 °C (up to 34 °C with limited temperature range) |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
Packing/shipping data
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | B x H x T: 1330 x 2150 x 1050 mm |
Net weight | approx. 249 kg |
Gross weight carton | approx. 319 kg |